VALLEYLAB® FORCE FX
Force FX: Digital Electrosurgical Unit
Generators
with computer controlled Instant Response™ Technology designed to
ensure a consistent clinical effect through all tissue types.
Specifications:
Dimensions: 4 3/8” H x 14” W x 17” L (11.1 x 35.6 x 43.2 cm)
Weight: 18 lbs. (8.2 kg)
Input power requirements: Operating range: 170 to 264 VAC. Maximum current: 3.5 A in cut and 2 A in coag.
Output Waveforms
Bipolar: Precise, standard and Macro: 470 kHz sinusoid
Monopolar Cut:
Low: 390 kHz sinusoid. Similar to the pure cut mode except the maximum voltage is limited to a lower value.
Pure: 390 kHz sinusoid.
Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals, 50% duty cycle envelope.
Monopolar Coag:
Desiccate: 240 kHz sinusoid repeated at 39 kHz, 8% duty cycle.
Fulgurate:
390 kHz damped sinusoidal bursts with a repetition frequency of 57 kHz
into 300 Ohm. Spray: 390 kHz damped sinusoid bursts with a randomized
repetition centered at 28 kHz. Frequencies include 21 kHz < f <35
kHz. Output is further modulated by a random 250 Hz envelope with a
variable duty cycle.
Low Frequency Leakage: 50-60
Hz. Source current, patient leads, all outputs tied together. Normal
polarity, intact chassis ground < 10μA. Normal polarity,
Ground open < 50μA. Reverse polarity, ground open < 50μA. Sink
Current at high line, all inputs < 50μA.
High Frequency Leakage: Bipolar: < 59.2 mArms. Monopolar: < 150 mArms. CEM™: < 150 mArms at < or equal 50 W.
Power Readouts: Agree with actual power into rated load to within ±15% or 5 watt, whichever is greater.